The BL625 is a high-precision, cost-effective single channel thermometer, designed to work with 25 Ω and 100 Ω platinum resistance thermometers (RTDs) or standard platinum resistance thermometers (SPRTs). The sensor parameters can be configured, supporting the ITS-90 temperature scale, CVD equation, and IEC60751 standard. It offers an accuracy of up to ±0.005°C and a full-scale resolution of 0.0001°C. Its unique design makes it suitable for a variety of applications, from lab calibration to industrial process measurements.
FEATURES
Supports measuring 25Ω and 100Ω platinum resistors(RTD, PRT, IPRT)
Supports measuring 25Ω and 100Ω SPRT
Supports ITS-90 characteristic coefficients
Supports CVD (Callendar Van Dusen) and IEC-60751 characteristic coefficients
Uses four-wire Kelvin connection to eliminate leadresistance effects
Uses constant current excitation: 1mA
Uses current reversal technology to eliminate thermoelectric EMF
High accuracy: ±0.005°C at 0°C
Resolution: 0.0001°C
Extremely low temperature coefficient: 0.3ppm/°C
Fast measurement: 1-second measurement cycle
Has filtering function, with configurable filter period
Units can be switched: Ω, °C, ℉, K
User-calibratable
Sensor parameter password protection
Built-in parameter simulation & calculation function
Serial RS-232 interface for remote reading of measurement data and controlling the thermometer
Easy sensor connection: supports both terminal posts and DIN plugs for different scenarios
Compact, lightweight, portable, and sturdy
APPLICATION
Secondary laboratory calibration
Laboratory, education, research testing
Industries: chemical, pharmaceuticals and cosmetics; food and beverage; petroleum industry; renewable energy applications
SPECIFICATION
Supported Probes: 25Ω or 100Ω SPRT/RTD
Temperature Range: -196℃ ~ 920℃
Temperature Resolution: 0.0001℃
Temperature Accuracy: ±0.005℃ (@0℃) *
Resistance Range: 5Ω ~ 400Ω
Resistance Resolution: 0.0001 Ω (5~ 20Ω), 0.001Ω (20~ 400Ω)
Resistance Accuracy: 0.0005 Ω (5~ 20Ω), 20 ppm (20~ 400Ω)
Interpolation Algorithms: ITS-90: Temperature regions 4,6,7,8,9,10 and 11; CVD: R0, A, C and C; IEC60751
Excitation Current: 1mA
Sampling Frequency: 1Hz
Filter Cycle: 0~60s adjustable
Communication Interface: RS-232 Serial Interface (Female)
Communication Protocol: SCPI compatible
Connectors: 4mm banana plug / terminal / LEMO 1B 4-pin
Operating Temperature Range: 20℃ ±5℃
Operating Humidity Range: <85% RH (non-condensing)
Power Supply: AC 100-240V (±5%), 50/60Hz **
Dimensions: 155mm (W) x 195mm (L) x 65mm (H)
Weight: 1kg
Package Dimensions: 25cm (W) x 30cm (L) x 15cm (H)
Package Total Weight: 1.6kg ***
Note:
* Temperature measurement accuracy and range are limited by the sensor'sprecision and temperature span
** Customizable voltage versions like 110V/115V/230V are available
*** Total weight includes the instrument, accessories, and packaging
SCOPE OF DELIVERY
Precision reference thermometer
Protection case
OPTIONAL ACCESSORIES AND SERVICCE
ISO17025 calibration service (PN: CAL-625)
Reference PRT
ORDERING MODEL #
BL625
The worldwide product accuracy comparison:
Fluke 1595A: Measurement range 0~500 kΩ, accuracy ±0.06 ~ ±0.2 ppm
Fluke 1594A: Measurement range 0~500 kΩ, accuracy ±0.8 ppm
ASL F18/F700/F700A: Measurement range 0~100 kΩ, accuracy ±0.1 ~ ±0.5 ppm
Isotech millik/ MicroK: Measurement range 1~100 kΩ, accuracy ±3 ~ ±5 ppm
Isotech TTI-22: Measurement range 0~440 Ω, accuracy ± 5 ppm
Baluntech BL621: Measurement range 5~3400 Ω, accuracy ±5 ppm (we are here!)
Fluke 1586A: Measurement range 0~4000 Ω, accuracy ±10 ~ ±20 ppm
Baluntech BL625: Measurement range 5~400 Ω, accuracy ±20 ppm (we are here!)
Fluke 1529: Measurement range 0~400 Ω, accuracy ± 25 ppm
Fluke 1502A: Measurement range 0~400 Ω, accuracy ± 25 ppm
Fluke 1523: Measurement range 0~400 Ω, accuracy ± 40 ppm +0.002 Ω
Fluke 1524: Measurement range 0~400 Ω, accuracy ± 40 ppm +0.002 Ω
Omega, Honeywell, Siemens, Eerson, E+H industry thermometers: ±50 ppm