BL789 Dew Point Chilled Mirror Hygrometer (-100℃ to 20℃ dp)

share
copy Link
Facebook
X
LinkedIn
Pinterest
Reading 0.01℃
Accuracy ±0.10℃ DP
Range (-100 to +20) ℃ dp
DescriptionFAQ?

       

The chilled mirror dew point meter is a precisetechnique for determining the water vaporcontent is gases by directly measuring dewpoint or frost temperatures. Using thistechnique a metal mirror is cooled until itreaches a temperature at which a thin layer ofcondensation begins to form on it. The dewlayer is detected optically, and the mirror isheld at that temperature. The mirrortemperature, measures with a PRT, is anaccurate indicator of dew or frost point.


KEY FEATURE

  • The dew point range is lower to -100℃

  • ±0.1℃ dew point accuracy

  • ±0.01℃ sensitivity and fast response

  • Real time curve and data logging

  • With multi-points calibration

  • Dynamic contamination correction

  • Manual Forced frost function

  • Auto process the super cooled water

  • Analog output: Optional




APPLICATION

  • Standard reference for calibration

  • Precision moisture measurements

  • Monitoring dried air pipe at a production

  • Moisture control at drying process

  • Production process of lithium ion battery

  • Measurement of gas with low dew point

  • Reference in humidity calibration facilities

  • Reference for RH sensor manufacturing




SPECIFICATION

  • Dew point measure range: (-70/-85/-100 to +20) ℃ dp

  • Dew point calibration range: (-60/-80/-90 to +20) ℃ dp

  • Dew point accuracy: ±0.1℃

  • Dew point resolution: 0.01℃

  • Dew point sensitivity: ±0.01℃

  • Dew point reproducibility: ±0.05℃

  • Response: Stable measure at +15°C dp within 1 minute

  • Flow range: 0-2 liter/minute

  • Flow accuracy: <= ±0.25%FS

  • Sample flow rate: 0.3 to 1 Nl/min (suggest 0.7 Nl/min)

  • Sampling gas pressure: < 1.5MPa (15barg)

  • Test gas: N2, H2, O2, CA, SF6, C2H4, C3H6, ...etc

  • Cooling: 4 Stage Peltier, air compressor

  • Display: 7 inch touch screen

  • Digital output: RS232

  • Gas in/out connector: 1/4 inch Swagelok tube

  • Display units: ℃ dp, ℃, %RH, PPMv, L/min

  • Power supply:100-240VAC

  • Power consume: 60W

  • Working condition: -20℃ to +50℃, < 95%RH

  • Dimension: W750mm x D680mm x H400mm

  • Weight: 12kg


SCOPE OF DELIVERY

  • BL789 Chilled mirror hygrometer

  • Standard AC Power cord

 

 

OPTIONAL

  • ISO17025 calibration certificate (Part #: CAL-789)

  • Dry gas generator (Part #: BL792)

  • Dew Point Generator (Part #: BL793)

  • Silent oil-free air compressor (buy locally)

  • Integrated cabinet with pipes, calibration bench, manifold (Part #: 790-PACK )

ORDERING MODEL #

  • BL789-60: ±0.1℃ accuracy from -60 ℃ to 20℃

  • BL789-80: ±0.1℃ accuracy from -80 ℃ to 20℃

  • BL789-90: ±0.1℃ accuracy from -90 ℃ to 20℃


Our cooling methods:

1. -60C DP: Semiconductor refrigeration

2. -80C DP: Compressor refrigeration

3. -90C DP: Stirling refrigeration (1~2W@80K, ≈ -193℃)


How to fast generate the dew point and short the stable time?

We have resolved the following three points:

1. Temperature acquisition data isstable

2. Mirror surface temperature control is stable

3. Light intensityattenuation is stable.

The relative algorithms for above three points can form a control technical closed loop.




Q: What is the lowest measurement dew point ?

Baluntech did -105℃ DP in our lab, as below:


Test  -102C DP.jpg