The chilled mirror dew point meter is a precisetechnique for determining the water vaporcontent is gases by directly measuring dewpoint or frost temperatures. Using thistechnique a metal mirror is cooled until itreaches a temperature at which a thin layer ofcondensation begins to form on it. The dewlayer is detected optically, and the mirror isheld at that temperature. The mirrortemperature, measures with a PRT, is anaccurate indicator of dew or frost point.
KEY FEATURE
The dew point range is lower to -100℃
±0.1℃ dew point accuracy
±0.01℃ sensitivity and fast response
Real time curve and data logging
With multi-points calibration
Dynamic contamination correction
Manual Forced frost function
Auto process the super cooled water
Analog output: Optional
APPLICATION
Standard reference for calibration
Precision moisture measurements
Monitoring dried air pipe at a production
Moisture control at drying process
Production process of lithium ion battery
Measurement of gas with low dew point
Reference in humidity calibration facilities
Reference for RH sensor manufacturing
SPECIFICATION
Dew point measure range: (-70/-85/-100 to +20) ℃ dp
Dew point calibration range: (-60/-80/-90 to +20) ℃ dp
Dew point accuracy: ±0.1℃
Dew point resolution: 0.01℃
Dew point sensitivity: ±0.01℃
Dew point reproducibility: ±0.05℃
Response: Stable measure at +15°C dp within 1 minute
Flow range: 0-2 liter/minute
Flow accuracy: <= ±0.25%FS
Sample flow rate: 0.3 to 1 Nl/min (suggest 0.7 Nl/min)
Sampling gas pressure: < 1.5MPa (15barg)
Test gas: N2, H2, O2, CA, SF6, C2H4, C3H6, ...etc
Cooling: 4 Stage Peltier, air compressor
Display: 7 inch touch screen
Digital output: RS232
Gas in/out connector: 1/4 inch Swagelok tube
Display units: ℃ dp, ℃, %RH, PPMv, L/min
Power supply:100-240VAC
Power consume: 60W
Working condition: -20℃ to +50℃, < 95%RH
Dimension: W750mm x D680mm x H400mm
Weight: 12kg
SCOPE OF DELIVERY
BL789 Chilled mirror hygrometer
Standard AC Power cord
OPTIONAL
ISO17025 calibration certificate (Part #: CAL-789)
Dry gas generator (Part #: BL792)
Dew Point Generator (Part #: BL793)
Silent oil-free air compressor (buy locally)
Integrated cabinet with pipes, calibration bench, manifold (Part #: 790-PACK )
ORDERING MODEL #
BL789-60: ±0.1℃ accuracy from -60 ℃ to 20℃
BL789-80: ±0.1℃ accuracy from -80 ℃ to 20℃
BL789-90: ±0.1℃ accuracy from -90 ℃ to 20℃
Our cooling methods:
1. -60C DP: Semiconductor refrigeration
2. -80C DP: Compressor refrigeration
3. -90C DP: Stirling refrigeration (1~2W@80K, ≈ -193℃)
How to fast generate the dew point and short the stable time?
We have resolved the following three points:
1. Temperature acquisition data isstable
2. Mirror surface temperature control is stable
3. Light intensityattenuation is stable.
The relative algorithms for above three points can form a control technical closed loop.
Q: What is the lowest measurement dew point ?
Baluntech did -105℃ DP in our lab, as below: