BL512 High-Pressure Deadweight Tester (800MPa)BL512 High-Pressure Deadweight Tester (800MPa)BL512 High-Pressure Deadweight Tester (800MPa)BL512 High-Pressure Deadweight Tester (800MPa)BL512 High-Pressure Deadweight Tester (800MPa)BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • 1/6
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)
  • BL512 High-Pressure Deadweight Tester (800MPa)

BL512 High-Pressure Deadweight Tester (800MPa)

share
copy Link
Facebook
X
LinkedIn
Pinterest
Accuracy ±0.05%RD
Range 800MPa (120,000psi)
DescriptionFAQ?

 

The model BL512 high pressure deadweight testers are applicable in very wide variety of pressure calibration and measurement tasks. The pressure ranges is up to 800MPa. The piston cylinder assembly is the heart of the dead weight tester and responsible for its high accuracy. We manufactures piston cylinder with accuracies from 0.05% up to 0.005% of reading. To give an idea of the manufacturing skills, the space between the piston and the cylinder is in the um range and can only be measured in a climate room.

   

       

KEY FEATURES

  • 5MPa to 500MPa,… 20MPa to 800MPa

  • 0.01%, 0.02%, 0.05% accuracy

  • With piston rotation electric motor

  • 3.5 inch LED displays piston location, ambient temperature and humidity

  • 400ml oil reservoir

  • 16mm thick aluminum alloy base

  • Piston position can be easily calibrated

  • Modulzied piston and cylinder unit

  • The piston and cylinder is made of the high intensity, extreme hardness and lowest thermal expansion tungsten carbide, which , density is 15 g/cm3, thermal coefficient of expansion is 16.5 ppm /℃

     

APPLICATION  

  • Pressure calibration labs

  • Pressure measurement in field

   




Youtube video links https://youtu.be/mBYArnE43QE


SPECIFICATION

  • Pressure range: (5~250)MPa, (20~800)MPa

  • Accuracy: 0.005%, 0.01%, 0.02%, 0.05%

  • Piston material: Tungsten carbide

  • Cylinder material: Tungsten carbide

  • Mass set material: 304 non-magnetic stainless steel

  • Hydraulic base material: Aluminium alloy

  • Transforming and lubrication medium: Sebacate

  • Piston diameter: 0.16cm

  • Effective area: 0.02cm2

  • Reservoir volume: 400ml

  • Connector: Metric (NPT and BSP is option)

  • Pressure units available: MPa, bar, PSI

  • Base size: 360 mm (H) x 710 mm (L) x 510 mm (D)

  • Base weight: 30kg (40kg for 500MPa and 600MPa)



Part #

Pressure

Piston

Mass

Accuracy

BL512-250MPA

(5~250) MPa

1kg

49kg

0.005

0.01

0.02

0.05

BL512-300MPA

(5~300) MPa

1kg

59kg

0.005

0.01

0.02

0.05

BL512-400MPA

(5~400) MPa

1kg

79kg

0.01

0.02

0.05

BL512-500MPA

(5~500) MPa

1kg

99kg

0.01

0.02

0.05

BL512-600MPA

(10~600) MPa

2kg

118kg

0.01

0.02

0.05

BL512-800MPA

(20~800) MPa

4kg

156kg

0.05


           

SCOPE OF DELIVERY

  • BL512 high pressure deadweight tester base

  • Piston cylinder unit

  • 304 non-magnetic stainless steel mass set

  • Mass carrier plate

  • Adapter kits (9/16 UNF, 1/2 BSP)

  • Transit case


OPTIONAL ACCESSORIES AND SERVICES

  • Accredited Calibration, Include mass set   (P/N: CAL-512)

 

ORDERING MODEL NUMBER


  • BL512-250MPA-X

  • BL512-300MPA-X

  • BL512-400MPA-X

  • BL512-500MPA-X

  • BL512-600MPA-X

  • BL512-800MPA-X (X is accuracy. 005: 0.005%; 01: 0.01%; 02: 0.02%; 05: 0.05%)        

  (need to inform the local gravity value in advance)


   



Q: What is the length of piston?

     Generally, it is 50mm with exception of inserting part of piston seat.

     

Q: What is the temperature on the front LED of BL512 deadweight tester?

      It is amibent temperature.

      The front LED also show the ambient humidity.

      The temperature and humidity sensor is located under the hydraulic base.

       

Q: How to measure the piston temperature of BL512 deadweight tester?

      Generally, use a AA class PT100, contact the outside of PCU to read the temperature value.

Q: There is a small hole on the piston cylinder unit, what is it used for?

      It is used to release the excess librucating oill of PCU system.

   


Q: The mass quantities for BL512-500MPA-X?

Pressure (MPa)Quantity (pcs)Diameter (mm)Mass (kg/pc)Material
25182805304SS
2012554304SS
1022002304SS
511601304SS