BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)
  • BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)
  • BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)
  • 1/2
  • BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)
  • BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)

BL782 Precision Chilled Mirror Hygrometer (-30 to 75℃ dp)

share
copy Link
Facebook
X
LinkedIn
Pinterest
Accuracy ±0.10℃ DP
Range (-30 to +75) ℃ dp
DescriptionFAQ?

                   

Chilled mirror dew point meter is a precise technique for determining the water vapor content is gases by directly measuring dew point or frost temperatures. Using this technique a metal mirror is cooled until it reaches a temperature at which a thin layer of condensation begins to form on it. The dew layer is detected optically, and the mirror is held at that temperature. The mirror temperature, measures with a PRT, is an accurate indicator of dew or frost point.



FEATURES

  • Accuracy of ±0.10 °C

  • Measurement range -30℃ to +75℃

  • ±0.01℃ sensitivity and fast response

  • Real time curve and data logging

  • With multi-points calibration

  • Dynamic contamination correction

  • Manual Forced frost function

  • Auto process the super cooled water



APPLICATION

  • Reference standard in calibration lab

  • Inlet air measurement for engine testing

  • Heating ventilation & air conditioning test

  • Environment monitoring in clean room

  • Wet control for metallurgical processes

  • Coating process monitoring in pharmacy

  • Reference humidity instruments for climatic and environmental chambers

  • Lithium Ion battery manufacture



SPECIFICATION

  • Dew point measurement range: (-40 to +90) ℃ dp

  • Dew point calibration range: (-25 to +70) ℃ dp

  • Dew point accuracy: ±0.10℃

  • Dew point resolution: 0.01℃

  • Dew point sensitivity: ±0.01℃

  • Dew point reproducibility: ±0.05℃

  • Response: Stable measure at +15°C dp within 1 minute

  • Temperature measurement range: -40℃ to +100℃

  • Temperature calibration range: -10℃ to +90℃

  • Temperature accuracy: ±0.05℃

  • Humidity range: 3.6...100%RH @23℃

  • Display units: ℃ dp, ℃, %RH, PPMv, L/min

  • Mirror material: Gold-plated copper

  • Mirror temperature sensor: Pt100, AAA (1/10 DIN)

  • Mirror Cooling: 3 Stage PELTIER

  • External Temperature sensor: Pt100, Class AA (1/3 DIN)

  • Cable material: Polyurethane with shielding layer

  • Sampling airflow pressure: < 1.5MPa

  • Display: 7 inch touch screen

  • Digital output: RS232

  • Analog output: Optional

  • Power supply:100-240VAC

  • Power consume: 12W

  • Ingress protection grade: IP54

  • Working condition: -20℃ to +50℃, < 95%RH

  • Dimension: W248mm x D240mm x H140mm

  • Weight: 3.5 kg



SCOPE OF DELIVERY

  • BL782 Chilled mirror hygrometer

  • Dew point sensor (M36 x 1.5 thread) with 2 meters cable

  • Ø3 x 115 mm Pt100 with 2 meters cable

  • Standard AC Power cord

  • Water-proof carrying case




OPTIONAL

  • ISO17025 accredited calibration certificate

  • Analog output 4-20mA


ORDERING MODEL #

  • BL782



         

What is the use of forced frost function?

Normally, the water vapor may be in the state of liquid or solid phase (dew or frost) if the temperature is below 0 ℃. The difference in the temperature at which the condensate layer stabilizes can be up to 3 ˚C, therefore the condensate phase must be known for correct calculation or validation of parameters such as relative humidity.

However, it is also possible that dew and frost can exist concurrently on the mirror; this results in a non-stable value somewhere between the dew and frost point.

Force frost function overcools the mirror to force the condensed layer to the solid phase. This eliminates the uncertainty of whether dew or frost point is measured.

After this function is enabled, the device will automatically trigger the forced frosting if the following conditions are met:

A. -35℃<DP<-3℃;

B. The sensor is in stable control state for 30 seconds;

C. Frost thickness is from 80% to 120%, and stable for 30 seconds.



How to caculate the relative humidity ?

**Appendix 1: Dew Point – Humidity Conversion**


This appendix provides the formula and results for converting between dew point of moisture in gas and µL/L (PPMv).


The conversion from dew point to µL/L (PPMv) is calculated using the following formula:

计算公式1.png …………………………(1)

Where:

K——Moisture content by volume, µL/L or PPMv;

Ps——Saturated vapor pressure of ice at the outlet temperature, Pa.


The saturated vapor pressure at system pressure is calculated using the following formula:

计算公式2.png………………………… (2)

Where:

Psvo——Saturated vapor pressure at system pressure, Pa;

Psva——Saturated vapor pressure at atmospheric pressure, Pa;

P0——System pressure, Pa;

Pa——Atmospheric pressure, Pa.


Derived from formulas (1) and (2), the conversion method for measuring moisture content under system pressure is:

计算公式3.png………………… (3)

Where:

K0——Moisture content under system pressure, µL/L or PPMv;

Ka——Moisture content displayed by the instrument, µL/L or PPMv.