BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • 1/7
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)
  • BL569 Compensated Micro Manometer (±2500Pa)

BL569 Compensated Micro Manometer (±2500Pa)

share
copy Link
Facebook
X
LinkedIn
Pinterest
Accuracy ± 0.4Pa
Range ± 2500 Pa
DescriptionFAQ?

The BL569   compensated micro manometer is a precision low pressure balance. It equipped a big movable reservoir and a small fixed reservoir. By moving up and down the big reservoir, the pressure generated by the water level difference between the big and small reservoir is balanced with the desired pressure from external pressure calibration pump.   It is very easy to calibrate low pressure, vacuum and differential pressure in laboratory.  

FEATURES

  • Ranges ± 2500 Pa

  • Accuracy is up to ±0.4Pa

  • The transparent observation tube is equipped with a water gauge head, which can observe the level changes in big reservoir.

  • The reflective mirror can show the contact between the water level head and the liquid level inside the observation tube.

  • Fine adjust disc with 200 equal divides is fixed on a long screw. The long screw moves big reservoir up and down. The bottom scales shows the height of big reservoir.

 

APPLICATION

  • Coal mine

  • Low pressure calibration in lab.

  • Low pressure test, vacuum test, differential pressure test (non corrosive gases)

 

VIDEO





YouTube video links: https://youtu.be/u4IqyNyeqys


   

SPECIFICATION

  • Pressure range: -2500Pa to 2500Pa (±10 inch Water)

  • Accuracy class: 0.02

  • Tolerance error:   ±0.4Pa (-1500 Pa to +1500 Pa);   ± 0.5Pa   (-2500 Pa to -1500 Pa) ; ± 0.5Pa (+1500 Pa to +2500 Pa)

  • Liquid level scale resolution: 0.01mm

  • Liquid in reservoirs: Distilled water, deionized water

  • Stability: Maintain at 2500Pa for 3 minutes, and no pressure drop in the next 1 minute

  • Maximum pressure: Maintain at 5000Pa for 10 minutes, and no pressure drop in the next 5 minutes

  • Zero alignment tolerated error:< ±0.1Pa

  • Zero return tolerated error: < ±0.2Pa

  • Dimension: 230mm x 322mm

  • Weight: 6kg

   

SCOPE OF DELIVERY

  • BL569 compensated micro manometer

  • Flexible connection hose x 2

  • User manual

  • Transit case

   

OPTIONAL ACCESSORIES AND SERVICES

  • ISO17025 calibration certificate (P/N: CAL-569)

  • Low pressure calibration pump

       

ORDERING MODEL NUMBER

  • BL569-2500-02

   


 


Working principle?

The working principle of BL569 compensated micro manometer   is to compensate for the decrease in water level caused by external pressure by raising the liquid level of the large container, so that the water level of the small container returns to its original zero position. That is, the compensation principle is used to balance the pressure generated by the liquid level difference between the large and small containers with the external input pressure.

BL569-Compensated-Micro-Manometer -Working Principal.jpg


How to calculate the water level "height " of big reseroir?

P‍lease download the calculator tool. ‍


How to check the water density in different temperature?

Please refer the follow web links:

https://www.internetchemistry.com/chemical-data/water-density-table.php



How much volume of water is needed in system?

about 65 ml