Chilled mirror dew point meter is a precise technique for determining the water vapor content is gases by directly measuring dew point or frost temperatures. Using this technique a metal mirror is cooled until it reaches a temperature at which a thin layer of condensation begins to form on it. The dew layer is detected optically, and the mirror is held at that temperature. The mirror temperature, measures with a PRT, is an accurate indicator of dew or frost point.
FEATURES
+95℃ ultra high dew point temperature
±0.10℃ dew point accuracy
±0.01℃ sensitivity and fast response
Real time curve and data logging
With multi-points calibration
Dynamic contamination correction
Manual forced frost function
Auto process the super cooled water
APPLICATION
Reference standard in calibration lab
Inlet air measurement for engine testing
Heating ventilation & air conditioning test
Environment monitoring in clean room
Wet control for metallurgical processes
Coating process monitoring in pharmacy
Reference humidity instruments for climatic and environmental chambers
SPECIFICATION
Dew point measurement range: (10 to 95) ℃ dp
Dew point calibration range: (10 to 90) ℃ dp
Dew point accuracy: ±0.10℃ DP
Dew point resolution: 0.01℃
Dew point sensitivity: ±0.01℃
Dew point reproducibility error: ±0.05℃
Response: Stable measure at +15°C dp within 1 minute
Temperature measurement range: (-20 to +150) ℃
Temperature calibration range: (-10 to +100) ℃
Temperature accuracy: ±0.1℃
Sampling airflow pressure: < 1.5MPa
Display units: ℃ dp, ℃, %RH, PPMv, L/min
Mirror temperature sensor: Pt100, AAA (1/10 DIN)
Mirror Cooling: 3-Stages PELTIER
External Temperature sensor: Pt100, Class AA (1/3 DIN)
Cable material: Polyurethane with shielding layer
Display: 7 inch touch screen
Digital output: RS232
Analog output: Optional
Power supply:100-240VAC
Ingress protection grade: IP54
Working condition: -20℃ to +50℃, < 95%RH
Dimension: W248mm x D240mm x H140mm
Weight: 3.5 kg
SCOPE OF DELIVERY
BL781 High temperature chilled mirror hygrometer
3-Stage dew point sensor with 2 meters cable
External temperature sensor with 2 meters cable
Standard AC Power cord
Heated hose and steam trap
Water-proof carrying case
OPTIONAL
ISO17025 accredited calibration certificate
Analog output 4-20mA
ORDERING MODEL #
BL781
How to caculate the relative humidity ?
**Appendix 1: Dew Point – Humidity Conversion**
This appendix provides the formula and results for converting between dew point of moisture in gas and µL/L (PPMv).
The conversion from dew point to µL/L (PPMv) is calculated using the following formula:
…………………………(1)
Where:
K——Moisture content by volume, µL/L or PPMv;
Ps——Saturated vapor pressure of ice at the outlet temperature, Pa.
The saturated vapor pressure at system pressure is calculated using the following formula:
………………………… (2)
Where:
Psvo——Saturated vapor pressure at system pressure, Pa;
Psva——Saturated vapor pressure at atmospheric pressure, Pa;
P0——System pressure, Pa;
Pa——Atmospheric pressure, Pa.
Derived from formulas (1) and (2), the conversion method for measuring moisture content under system pressure is:
………………… (3)
Where:
K0——Moisture content under system pressure, µL/L or PPMv;
Ka——Moisture content displayed by the instrument, µL/L or PPMv